Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 741: (1 year) CAL VER /5520,3458,6666,2189 INSTRUMENT: Fluke 743: (1 year) CAL VER /5520,3458,6666,2189 INSTRUMENT: Honeywell 2020: (1 year) CAL VER /5520,3458,6666,2189 DATE: 27-Oct-98 AUTHOR: Fluke REVISION: $Revision: 1.3 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 103 NUMBER OF LINES: 1152 CONFIGURATION: Fluke 5520A CONFIGURATION: HP 3458A CONFIGURATION: Philips PM 6666 STANDARD: Fluke 2189A ============================================================================= # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # Source: # Fluke 74X Series Calibration Manual PN 602505 (February 1997). # # Compatibility: # MET/CAL 5.0 or later # # Subprocedures: # None # # Required Files: # 52_70x_V.bmp Picture: 5520A/74X Volts Measurement connection # 52_70xCW.bmp Picture: 5520A/74X Ohms/2-Wire RTD Measurement connection # 52_70xR3.bmp Picture: 5520A/74X 3-Wire RTD Measurement connection # 52_70xR4.bmp Picture: 5520A/74X 4-Wire RTD Measurement connection # 52_70xTC.bmp Picture: 5520A/74X TC Measurement/Source connection # 52_70x_A.bmp Picture: 5520A/74X Amps Measurement connection # 58_70xST.bmp Picture: 5520A,3458A/74X Simulate Transmitter connection # 58_70x_V.bmp Picture: 3458A/74X Volts Source connection # 58_70x_A.bmp Picture: 3458A/74X Amps Source connection # 58_70xLP.bmp Picture: 3458A/74X Loop Power connection # 58_70xCW.bmp Picture: 3458A/74X Ohms/2-Wire RTD Source connection # 58_74X_A.bmp Picture: DC Current Measurement connection # 66_70xfr.bmp Picture: PM6666/74X Freq/Duty Cycle connection # metlink3.exe Executable: Fluke 741/743 Serial Communication Program # ini_val.exe Executable: Gets default port for UUT from METCAL.INI # # Register usage by DOS program "METLINK3": # MEM UUT reading # MEM1 Error flag # 1 no error or error, ignore # 0 error, retry (repeat step) # -1 error, abort procedure # # METLINK3 usage: # # DC Voltage Measurement Verification # METLINK3 S[etup] M[easurement] DCV # METLINK3 R[ead] DCV # # AC Voltage Measurement Verification # METLINK3 S[etup] M[easurement] A[CV] # METLINK3 R[ead] A[CV] # # Frequency Measurement Verification # METLINK3 S[etup] M[easurement] F[req] # METLINK3 R[ead] F[req] # # Resistance Measurement Verification # METLINK3 S[etup] M[easurement] O[hms] # METLINK3 R[ead] O[hms] # # 2-Wire RTD Measurement Verification # METLINK3 S[etup] M[easurement] R[TD] 2[-Wire] [[degC] [PT385]] # METLINK3 R[ead] R[TD] # # 3-Wire RTD Measurement Verification # METLINK3 S[etup] M[easurement] R[TD] 3[-Wire] [[degC] [PT385]] # METLINK3 R[ead] R[TD] # # 4-Wire RTD Measurement Verification # METLINK3 S[etup] M[easurement] R[TD] 4[-Wire] [[degC] [PT385]] # METLINK3 R[ead] R[TD] # # TC Measurement Verification # METLINK3 S[etup] M[easurement] T[C] [degC [K]] # METLINK3 R[ead] T[C] # # TC Source Verification # METLINK3 S[etup] S[ource] T[C] [degC [K]] # # DC Current Measurement Verification # METLINK3 S[etup] M[easurement] DCI # # Simulate Transmitter Verification # METLINK3 S[etup] S[ource] S[imulate_Transmitter] # # DC Current Source Verification # METLINK3 S[etup] S[ource] DCI # # Frequency Source Verification # METLINK3 S[etup] S[ource] F[req] # # DC Voltage Source Verification # METLINK3 S[etup] S[ource] DCV # # Resistance Source Verification # METLINK3 S[etup] S[ource] O[hms] # # RTD Source Verification # METLINK3 S[etup] S[ource] R[TD] [degC [PT385]] # # Loop Power Verification # METLINK3 S[etup] S[ource] L[oop_pwr] 24 | 28 # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R Q N P F V 1.002 STD Fluke 2189A 1.003 HEAD INITIAL OPERATIONS 1.004 OPBR Has an "ACAL ALL" been performed on the system 1.004 OPBR 3458A multimeter in the last 24 hours? 1.005 MATH M[1] = MEM1 1.006 JMPT 1.008 1.007 IEEE [@3458]END ON; ACAL ALL[GTL] 1.008 OPBR Has a "5520A ZERO" been performed on the system 1.008 OPBR 5520A in the last 24 hours? 1.009 JMPT 1.014 1.010 DISP The 5520A's published specifications are contingent 1.010 DISP on performing a "ZERO" operation within 24 hours of 1.010 DISP use. Therefore, this function is performed prior to 1.010 DISP proceeding with the UUT's verification. 1.010 DISP 1.010 DISP (The 5520A ZERO operation takes approximately 1.010 DISP 9 minutes to complete.) 1.011 DISP Connect a polarized COPPER short to the 1.011 DISP 5520A TC terminals. 1.011 DISP 1.011 DISP (Ensure all other 5520A terminals are open, 1.011 DISP prior to proceeding.) 1.012 HEAD Performing 5520A ZERO, Please Wait... 1.013 IEEE [@5520][T0] CAL_START ZERO; CAL_NEXT; *OPC?[I] 1.014 MATH MEM1 = M[1] 1.015 JMPT 1.017 1.016 DISP When the "ACAL ALL" function of the system 1.016 DISP 3458A is completed, (as evidenced by 1.016 DISP its display) press "Advance" to continue. 1.017 DISP Select the "TO LO" position of the 3458A GUARD 1.017 DISP switch (depressed). 1.017 DISP 1.017 DISP The 3458A GUARD switch should remain depressed 1.017 DISP throughout the remainder of this procedure. 1.018 MATH M[1] = 0 1.019 HEAD INITIAL CONDITIONS AND EQUIPMENT SETUP 1.020 DISP WARNING: 1.020 DISP HIGH VOLTAGE is used or exposed during the execution 1.020 DISP of this procedure. DEATH ON CONTACT may result if 1.020 DISP personnel fail to observe safety precautions. 1.021 DISP Turn on the UUT. 1.022 DISP Warm-up time requirement: 5 minutes 1.022 DISP Ambient temperature: 18C - 28C. 1.022 DISP Relative humidity: less than 90%. 1.023 DOS ini_val metcal.ini startup port 1.024 MATH MEM1 = ZCMPI(MEM2, "COM5520") 1.025 JMPZ 1.028 1.026 DISP The 5520A UUT Serial Port (COM5520) is not supported 1.026 DISP by the METLINK3 communication program. 1.026 DISP 1.026 DISP You must exit MET/CAL and change "port=COM5520" in 1.026 DISP METCAL.INI to an available PC port. 1.027 JMP 103.003 1.028 DISP Connect the UUT to [MEM2] using a Fluke Service Cable 1.028 DISP Part #949446. 1.029 DISP Connect a "K" type thermocouple to the 5520A 1.029 DISP TC connector observing polarity. The "K" type 1.029 DISP thermocouple and a Fluke 2189A both must be 1.029 DISP simultaneously immersed in a common lag bath 1.029 DISP and allowed to stabilize. 1.030 HEAD Test UUT For Rev. 2.0 or Greater 1.031 DOS METLINK3 FirmRev 1.032 JMPF 103.003 1.033 JMPZ 1.028 1.034 MATH MEM1 = MEM > 0 1.035 JMPT 1.037 1.036 JMP 1.042 1.037 MATH MEM1 = MEM >= 2.0 1.038 JMPT 1.042 1.039 JMP 1.040 1.040 HEAD PRELIMINARY OPERATIONS[7][7] 1.041 DISP UUT Firmware Rev is V[MEM] 1.041 DISP Update Firmware to appropriate revision before shipping. 1.042 HEAD PRELIMINARY OPERATIONS 1.043 DISP Press the UUT "SETUP" key. 1.043 DISP Press "NEXT PAGE" key two (2) times. 1.043 DISP Press "DOWN" arrow key two (2) times. Press "ENTER" 1.043 DISP Set Temperature Scale to "ITS-90". Press "ENTER". 1.043 DISP Press "DONE". 1.044 DISP Select the UUT MEASURE/SOURCE "split" screen mode using 1.044 DISP the MEAS/SOURCE button. 1.045 HEAD {DC VOLTAGE MEASUREMENT TEST}[7][D100][7][D100][7] 1.046 PIC 55_70x_V 1.047 HEAD DC VOLTAGE MEASUREMENT TEST {110mV Range} 1.048 DOS METLINK3 Setup Measure DCV 100E-3 1.049 JMPZ 1.019 1.050 JMPF 103.001 1.051 5520 0.000mV S 2W 1.052 DOS METLINK3 Read DCV 1.053 JMPZ 1.046 1.054 JMPF 103.001 1.055 MATH MEM1 = 0 1.056 MATH MEM = MEM * 1000 1.057 MEME 1.058 MEMC 110 mV 0.025% 0.015/ 2.001 5520 100.000mV S 2W 2.002 DOS METLINK3 Read DCV 2.003 JMPZ 2.001 2.004 JMPF 103.001 2.005 MATH MEM1 = 100 2.006 MATH MEM = MEM * 1000 2.007 MEME 2.008 MEMC 110 mV 0.025% 0.015/ 3.001 5520 -100.000mV S 2W 3.002 DOS METLINK3 Read DCV 3.003 JMPZ 3.001 3.004 JMPF 103.001 3.005 MATH MEM1 = -100 3.006 MATH MEM = MEM * 1000 3.007 MEME 3.008 MEMC 110 mV 0.025% 0.015/ 4.001 HEAD DC VOLTAGE MEASUREMENT TEST {1.1V Range} 4.002 DOS METLINK3 Setup Measure DCV 1 4.003 JMPZ 4.001 4.004 JMPF 103.001 4.005 5520 0.00000V S 2W 4.006 DOS METLINK3 Read DCV 4.007 JMPZ 4.001 4.008 JMPF 103.001 4.009 MATH MEM1 = 0 4.010 MEME 4.011 MEMC 1.1 V 0.025% 0.005/ 5.001 5520 1.00000V S 2W 5.002 DOS METLINK3 Read DCV 5.003 JMPZ 5.001 5.004 JMPF 103.001 5.005 MATH MEM1 = 1 5.006 MEME 5.007 MEMC 1.1 V 0.025% 0.005/ 6.001 5520 -1.00000V S 2W 6.002 DOS METLINK3 Read DCV 6.003 JMPZ 6.001 6.004 JMPF 103.001 6.005 MATH MEM1 = -1 6.006 MEME 6.007 MEMC 1.1 V 0.025% 0.005/ 7.001 HEAD DC VOLTAGE MEASUREMENT TEST {11V Range} 7.002 DOS METLINK3 Setup Measurement DCV 10 7.003 JMPZ 7.001 7.004 JMPF 103.001 7.005 5520 0.0000V S 2W 7.006 DOS METLINK3 Read DCV 7.007 JMPZ 7.001 7.008 JMPF 103.001 7.009 MATH MEM1 = 0 7.010 MEME 7.011 MEMC 11 V 0.025% 0.005/ 8.001 5520 10.0000V S 2W 8.002 DOS METLINK3 Read DCV 8.003 JMPZ 8.001 8.004 JMPF 103.001 8.005 MATH MEM1 = 10 8.006 MEME 8.007 MEMC 11 V 0.025% 0.005/ 9.001 5520 -10.0000V S 2W 9.002 DOS METLINK3 Read DCV 9.003 JMPZ 9.001 9.004 JMPF 103.001 9.005 MATH MEM1 = -10 9.006 MEME 9.007 MEMC 11 V 0.025% 0.005/ 10.001 HEAD DC VOLTAGE MEASUREMENT TEST {110V Range} 10.002 DOS METLINK3 Setup Measurement DCV 100 10.003 JMPZ 10.001 10.004 JMPF 103.001 10.005 5520 0.000V S 2W 10.006 DOS METLINK3 Read DCV 10.007 JMPZ 10.001 10.008 JMPF 103.001 10.009 MATH MEM1 = 0 10.010 MEME 10.011 MEMC 110 V 0.05% 0.005/ 11.001 5520 100.000V S 2W 11.002 DOS METLINK3 Read DCV 11.003 JMPZ 11.001 11.004 JMPF 103.001 11.005 MATH MEM1 = 100 11.006 MEME 11.007 MEMC 110 V 0.05% 0.005/ 12.001 5520 -100.000V S 2W 12.002 DOS METLINK3 Read DCV 12.003 JMPZ 12.001 12.004 JMPF 103.001 12.005 MATH MEM1 = -100 12.006 MEME 12.007 MEMC 110 V 0.05% 0.005/ 13.001 HEAD DC VOLTAGE MEASUREMENT TEST {300V Range} 13.002 DOS METLINK3 Setup Measurement DCV 300 13.003 JMPZ 13.001 13.004 JMPF 103.001 13.005 5520 0.00V S 2W 13.006 DOS METLINK3 Read DCV 13.007 JMPZ 13.001 13.008 JMPF 103.001 13.009 MATH MEM1 = 0 13.010 MEME 13.011 MEMC 300 V 0.05% 0.005/ 14.001 5520 300.00V S 2W 14.002 DOS METLINK3 Read DCV 14.003 JMPZ 14.001 14.004 JMPF 103.001 14.005 MATH MEM1 = 300 14.006 MEME 14.007 MEMC 300 V 0.05% 0.005/ 15.001 5520 -300.00V S 2W 15.002 DOS METLINK3 Read DCV 15.003 JMPZ 15.001 15.004 JMPF 103.001 15.005 MATH MEM1 = -300 15.006 MEME 15.007 MEMC 300 V 0.05% 0.005/ 16.001 5520 * S 16.002 HEAD {AC VOLTAGE MEASUREMENT TEST} 16.003 HEAD AC VOLTAGE MEASUREMENT TEST {1.1V Range} 16.004 DOS METLINK3 Setup Measurement ACV 1 16.005 JMPZ 16.001 16.006 JMPF 103.001 16.007 5520 0.2000V 500H SI S 2W 16.008 DOS METLINK3 Read ACV 16.009 JMPZ 16.001 16.010 JMPF 103.001 16.011 MATH MEM1 = 0.2 16.012 MEME 16.013 MEMC 1.1 V 0.5% 0.0005U 500H 17.001 5520 1.0000V 20H SI S 2W 17.002 DOS METLINK3 Read ACV 17.003 JMPZ 17.001 17.004 JMPF 103.001 17.005 MATH MEM1 = 1 17.006 MEME 17.007 MEMC 1.1 V 2% 0.0010U 20H 18.001 5520 1.0000V 40H SI S 2W 18.002 DOS METLINK3 Read ACV 18.003 JMPZ 18.001 18.004 JMPF 103.001 18.005 MATH MEM1 = 1 18.006 MEME 18.007 MEMC 1.1 V 0.5% 0.0005U 40H 19.001 5520 1.0000V 500H SI S 2W 19.002 DOS METLINK3 Read ACV 19.003 JMPZ 19.001 19.004 JMPF 103.001 19.005 MATH MEM1 = 1 19.006 MEME 19.007 MEMC 1.1 V 0.5% 0.0005U 500H 20.001 5520 1.0000V 1kH SI S 2W 20.002 DOS METLINK3 Read ACV 20.003 JMPZ 20.001 20.004 JMPF 103.001 20.005 MATH MEM1 = 1 20.006 MEME 20.007 MEMC 1.1 V 2% 0.0010U 1kH 21.001 5520 1.0000V 5kH SI S 2W 21.002 DOS METLINK3 Read ACV 21.003 JMPZ 21.001 21.004 JMPF 103.001 21.005 MATH MEM1 = 1 21.006 MEME 21.007 MEMC 1.1 V 10% 0.0020U 5kH 22.001 HEAD AC VOLTAGE MEASUREMENT TEST {11V Range} 22.002 DOS METLINK3 Setup Measurement ACV 10 22.003 JMPZ 22.001 22.004 JMPF 103.001 22.005 5520 2.000V 500H SI S 2W 22.006 DOS METLINK3 Read ACV 22.007 JMPZ 22.001 22.008 JMPF 103.001 22.009 MATH MEM1 = 2 22.010 MEME 22.011 MEMC 11 V 0.5% 0.005U 500H 23.001 5520 10.000V 500H SI S 2W 23.002 DOS METLINK3 Read ACV 23.003 JMPZ 23.001 23.004 JMPF 103.001 23.005 MATH MEM1 = 10 23.006 MEME 23.007 MEMC 11 V 0.5% 0.005U 500H 24.001 5520 10.000V 5kH SI S 2W 24.002 DOS METLINK3 Read ACV 24.003 JMPZ 24.001 24.004 JMPF 103.001 24.005 MATH MEM1 = 10 24.006 MEME 24.007 MEMC 11 V 10% 0.020U 5kH 25.001 HEAD AC VOLTAGE MEASUREMENT TEST {110V Range} 25.002 DOS METLINK3 Setup Measurement ACV 100 25.003 JMPZ 25.001 25.004 JMPF 103.001 25.005 5520 20.00V 500H SI S 2W 25.006 DOS METLINK3 Read ACV 25.007 JMPZ 25.001 25.008 JMPF 103.001 25.009 MATH MEM1 = 20 25.010 MEME 25.011 MEMC 110 V 0.5% 0.05U 500H 26.001 5520 100.00V 500H SI S 2W 26.002 DOS METLINK3 Read ACV 26.003 JMPZ 26.001 26.004 JMPF 103.001 26.005 MATH MEM1 = 100 26.006 MEME 26.007 MEMC 110 V 0.5% 0.05U 500H 27.001 5520 100.00V 5kH SI S 2W 27.002 DOS METLINK3 Read ACV 27.003 JMPZ 27.001 27.004 JMPF 103.001 27.005 MATH MEM1 = 100 27.006 MEME 27.007 MEMC 110 V 10% 0.20U 5kH 28.001 HEAD AC VOLTAGE MEASUREMENT TEST {300V Range} 28.002 DOS METLINK3 Setup Measurement ACV 300 28.003 JMPZ 28.001 28.004 JMPF 103.001 28.005 5520 50.0V 500H SI S 2W 28.006 DOS METLINK3 Read ACV 28.007 JMPZ 28.005 28.008 JMPF 103.001 28.009 MATH MEM1 = 50 28.010 MEME 28.011 MEMC 300 V 0.5% 0.5U 500H 29.001 5520 120.0V 60H SI S 2W 29.002 DOS METLINK3 Read ACV 29.003 JMPZ 29.001 29.004 JMPF 103.001 29.005 MATH MEM1 = 120 29.006 MEME 29.007 MEMC 300 V 0.5% 0.5U 60H 30.001 5520 250.0V 500H SI S 2W 30.002 DOS METLINK3 Read ACV 30.003 JMPZ 30.001 30.004 JMPF 103.001 30.005 MATH MEM1 = 250 30.006 MEME 30.007 MEMC 300 V 0.5% 0.5U 500H 31.001 5520 219.0V 5kH SI S 2W 31.002 DOS METLINK3 Read ACV 31.003 JMPZ 31.001 31.004 JMPF 103.001 31.005 MATH MEM1 = 219.0 31.006 MEME 31.007 MEMC 300 V 10% 2.0U 5kH 32.001 5520 * S 32.002 HEAD {FREQUENCY MEASUREMENT TEST} [7][7][7] 32.003 DISP Disconnect the jumpers from the three 'lows.' 32.004 DOS METLINK3 Setup Measurement Freq 10 32.005 JMPZ 32.001 32.006 JMPF 103.001 32.007 5520 10.00H 110mV SI S 2W 32.008 DOS METLINK3 Read Freq 32.009 JMPZ 32.001 32.010 JMPF 103.001 32.011 MATH MEM1 = 10 32.012 MEME 32.013 MEMC H 0.05U 110mV 33.001 DOS METLINK3 Setup Measurement Freq 500 33.002 JMPZ 33.001 33.003 JMPF 103.001 33.004 5520 500.0H 110mV SI S 2W 33.005 DOS METLINK3 Read Freq 33.006 JMPZ 33.001 33.007 JMPF 103.001 33.008 MATH MEM1 = 500 33.009 MEME 33.010 MEMC H 0.5U 110mV 34.001 DOS METLINK3 Setup Measurement Freq 10E+3 34.002 JMPZ 34.001 34.003 JMPF 103.001 34.004 5520 10.000kH 700mV SI S 2W 34.005 DOS METLINK3 Read Freq 34.006 JMPZ 35.001 34.007 JMPF 103.001 34.008 MATH MEM1 = 10 34.009 MATH MEM = MEM / 1000 34.010 MEME 34.011 MEMC kH 0.005U 700mV 35.001 DOS METLINK3 Setup Measurement Freq 50E+3 35.002 JMPZ 35.001 35.003 JMPF 103.001 35.004 5520 50.00kH 1.4V SI S 2W 35.005 DOS METLINK3 Read Freq 35.006 JMPZ 35.001 35.007 JMPF 103.001 35.008 MATH MEM1 = 50 35.009 MATH MEM = MEM / 1000 35.010 MEME 35.011 MEMC kH 0.05U 1.4V 36.001 5520 * S 36.002 HEAD {RESISTANCE MEASUREMENT TEST}[7][D100][7][D100][7] 36.003 PIC 55_70xCW 36.004 HEAD RESISTANCE MEASUREMENT TEST {11 Ohm Range} 36.005 DOS METLINK3 Setup Measurement Ohms 10 36.006 JMPZ 36.001 36.007 JMPF 103.001 36.008 5520 0.000Z S CW 36.009 DOS METLINK3 Read Ohms 36.010 JMPZ 36.001 36.011 JMPF 103.001 36.012 MATH MEM1 = 0 36.013 MEME 36.014 MEMC 11 Z 0.05% 0.050U 37.001 5520 10.000Z S CW 37.002 DOS METLINK3 Read Ohms 37.003 JMPZ 37.001 37.004 JMPF 103.001 37.005 MATH MEM1 = 10 37.006 MEME 37.007 MEMC 11 Z 0.05% 0.050U 38.001 HEAD RESISTANCE MEASUREMENT: {110 Ohm Range} 38.002 DOS METLINK3 Setup Measurement Ohms 100 38.003 JMPZ 38.001 38.004 JMPF 103.001 38.005 5520 0.00Z S CW 38.006 DOS METLINK3 Read Ohms 38.007 JMPZ 38.001 38.008 JMPF 103.001 38.009 MATH MEM1 = 0 38.010 MEME 38.011 MEMC 110 Z 0.05% 0.050U 39.001 5520 100.00Z S CW 39.002 DOS METLINK3 Read Ohms 39.003 JMPZ 39.001 39.004 JMPF 103.001 39.005 MATH MEM1 = 100 39.006 MEME 39.007 MEMC 110 Z 0.05% 0.05U 40.001 HEAD RESISTANCE MEASUREMENT: {1100 Ohm Range} 40.002 DOS METLINK3 Setup Measurement Ohms 1000 40.003 JMPZ 40.001 40.004 JMPF 103.001 40.005 5520 0.0Z S CW 40.006 DOS METLINK3 Read Ohms 40.007 JMPZ 40.001 40.008 JMPF 103.001 40.009 MATH MEM1 = 0 40.010 MEME 40.011 MEMC 1100 Z 0.05% 0.5U 41.001 5520 1000.0Z S CW 41.002 DOS METLINK3 Read Ohms 41.003 JMPZ 41.001 41.004 JMPF 103.001 41.005 MATH MEM1 = 1000 41.006 MEME 41.007 MEMC 1100 Z 0.05% 0.5U 42.001 HEAD RESISTANCE MEASUREMENT TEST {11 kOhm Range} 42.002 DOS METLINK3 Setup Measurement Ohms 10E+3 42.003 JMPZ 42.001 42.004 JMPF 103.001 42.005 5520 0.000kZ S CW 42.006 DOS METLINK3 Read Ohms 42.007 JMPZ 42.001 42.008 JMPF 103.001 42.009 MATH MEM1 = 0 42.010 MATH MEM = MEM / 1000 42.011 MEME 42.012 MEMC 11 kZ 0.1% 0.010U 43.001 5520 10.000kZ S CW 43.002 DOS METLINK3 Read Ohms 43.003 JMPZ 43.001 43.004 JMPF 103.001 43.005 MATH MEM1 = 10 43.006 MATH MEM = MEM / 1000 43.007 MEME 43.008 MEMC 11 kZ 0.1% 0.010U 44.001 5520 * S 44.002 HEAD {RTD MEASURE, 3-WIRE}[7][D100][7][D100][7][D100][7] 44.003 IEEE [@5520]TEMP_STD ITS_90 [D2000] 44.004 PIC 55_70xR3 44.005 DOS METLINK3 Setup Measurement RTD 3-Wire degC PT385 44.006 JMPZ 44.001 44.007 JMPF 103.001 44.008 5520 -180.0degC R1 S 3W 44.009 DOS METLINK3 Read RTD 44.010 JMPZ 44.001 44.011 JMPF 103.001 44.012 MATH MEM1 = -180 44.013 MEME 44.014 MEMC degC 0.7U 45.001 5520 100.0degC R1 S 3W 45.002 DOS METLINK3 Read RTD 45.003 JMPZ 45.001 45.004 JMPF 103.001 45.005 MATH MEM1 = 100 45.006 MEME 45.007 MEMC degC 0.9U 46.001 5520 780.0degC R1 S 3W 46.002 DOS METLINK3 Read RTD 46.003 JMPZ 46.001 46.004 JMPF 103.001 46.005 MATH MEM1 = 780 46.006 MEME 46.007 MEMC degC 1.2U 47.001 5520 * S 47.002 HEAD {RTD MEASURE, 4-WIRE}[7][D100][7][D100][7][D100][7] 47.003 PIC 55_70xR4 47.004 DOS METLINK3 Setup Measurement RTD 4-Wire degC PT385 47.005 JMPZ 47.001 47.006 JMPF 103.001 47.007 5520 -180.0degC R1 S 4W 47.008 DOS METLINK3 Read RTD 47.009 JMPZ 47.001 47.010 JMPF 103.001 47.011 MATH MEM1 = -180 47.012 MEME 47.013 MEMC degC 0.3U 48.001 5520 100.0degC R1 S 4W 48.002 DOS METLINK3 Read RTD 48.003 JMPZ 48.001 48.004 JMPF 103.001 48.005 MATH MEM1 = 100 48.006 MEME 48.007 MEMC degC 0.5U 49.001 5520 780.0degC R1 S 4W 49.002 DOS METLINK3 Read RTD 49.003 JMPZ 49.001 49.004 JMPF 103.001 49.005 MATH MEM1 = 780 49.006 MEME 49.007 MEMC degC 0.8U 50.001 5520 * S # # The system uncertainty for the following measurements was # determined using the report of calibration for the Fluke # 2189A, the report of calibration for the Omega "K" type # thermocouple and the linearity error for the Fluke 5520A. # The values used to calculate system uncertainty are as follows # Fluke 2189 uncertainty = 0.05 degC. # Omega "K" type thermocouple = 0.03 degC. # Linearity error of the Fluke 5520A = 0.33uV. # 50.002 HEAD {THERMOCOUPLE MEASUREMENT TEST} 50.003 HEAD THERMOCOUPLE MEASUREMENT TEST [7][D100][7][D100][7] 50.004 DISP The following steps are accomplished using the (ITS-90) 50.004 DISP millivolt realizations for "K" type thermocouples, as 50.004 DISP taken from the NIST Monograph 175. 50.005 MEMI Use keyboard to enter Fluke 2189A measured temperature. # M[17] = The Fluke 2189A measured value. 50.006 MATH MEM = FMT(MEM,"%3.2f") 50.007 MATH M[17] = MEM # M[18] = The value measured by Fluke 5520A (lag bath temperature). 50.008 IEEE [@5520] TC_MEAS CEL; TC_TYPE_D K; *OPC? [I]; VAL? [I$] 50.009 MATH MEM = FLD(MEM2,1,",") 50.010 MATH MEM = FMT(MEM2,"%3.2f") 50.011 MATH M[18] = MEM # M[19] = The ERROR (difference) expressed in degC. 50.012 MATH M[19] = (M[17] - M[18]) # M[20] = The CORRECTION (mV conversion) expressed in degC @ -180degC. 50.013 MATH M[20] = (M[19] * 2.153) # M[21] = The CORRECTION (mV conversion) expressed in degC @ 0degC. 50.014 MATH M[21] = (M[19] * 1.025) # M[22] = The CORRECTION (mV conversion) expressed in degC @ 1300degC. 50.015 MATH M[22] = (M[19] * 1.158) # M[23] = The CORRECTION (mV conversion) expressed in degC @ -200degC. 50.016 MATH M[23] = (M[19] * 2.651) # M[24] = The CORRECTION (mV conversion) expressed in degC @ 1370degC. 50.017 MATH M[24] = (M[19] * 1.193) 50.018 HEAD THERMOCOUPLE MEASUREMENT TEST 50.019 PIC 55_70xTC 50.020 HEAD Thermal Dissipation Delay, Please Wait 2 Min.[D30000] 50.021 HEAD Thermal Dissipation Delay, Please Wait 90 Sec.[D30000] 50.022 HEAD Thermal Dissipation Delay, Please Wait 1 Min.[D30000] 50.023 HEAD Thermal Dissipation Delay, Please Wait 30 Sec.[D30000] 50.024 HEAD THERMOCOUPLE MEASUREMENT TEST 50.025 DOS METLINK3 Setup Measurement TC degC K 50.026 JMPZ 50.001 50.027 JMPF 103.001 50.028 ACC -180.0degC 0.17117U 50.029 5520 -180degC _K TC S 2W 50.030 DOS METLINK3 Read TC 50.031 JMPZ 50.001 50.032 JMPF 103.001 50.033 MATH MEM1 = -180 + (M[20]) 50.034 MEME 50.035 MEMC degC 0.9U 51.001 ACC 0.0degC 0.08234U 51.002 5520 0.0degC _K TC S 2W 51.003 DOS METLINK3 Read TC 51.004 JMPZ 51.002 51.005 JMPF 103.001 51.006 MATH MEM1 = 0 + (M[21]) 51.007 MEME 51.008 MEMC degC 0.5U 52.001 ACC 800.0degC 0.07932U 52.002 5520 800.0degC _K TC S 2W 52.003 DOS METLINK3 Read TC 52.004 JMPZ 52.002 52.005 JMPF 103.001 52.006 MATH MEM1 = 800 + (M[19]) 52.007 MEME 52.008 MEMC degC 0.7U 53.001 ACC 1300.0degC 0.09292U 53.002 5520 1300.0degC _K TC S 2W 53.003 DOS METLINK3 Read TC 53.004 JMPZ 53.002 53.005 JMPF 103.001 53.006 MATH MEM1 = 1300 + (M[22]) 53.007 MEME 53.008 MEMC degC 0.9U 54.001 5520 * S 54.002 ASK- N 54.003 HEAD {THERMOCOUPLE SOURCE TEST} 54.004 HEAD THERMOCOUPLE SOURCE TEST 54.005 DOS METLINK3 Setup Source TC -200 degC K 54.006 JMPZ 54.001 54.007 JMPF 103.001 54.008 ACC -200.0degC 0.21682U 54.009 5520 -200.0degC _K TM N 2W 54.010 MATH MEM = MEM + (M[23]) 54.011 MATH MEM1 = -200 54.012 MEME 54.013 MEMC degC 0.6U 55.001 DOS METLINK3 Setup Source TC 0 degC K 55.002 JMPZ 55.001 55.003 JMPF 103.001 55.004 ACC 0.0degC 0.08234U 55.005 5520 0.0degC _K TM N 2W 55.006 MATH MEM = MEM + (M[21]) 55.007 MATH MEM1 = 0 55.008 MEME 55.009 MEMC degC 0.5U 56.001 DOS METLINK3 Setup Source TC 800 degC K 56.002 JMPZ 56.001 56.003 JMPF 103.001 56.004 ACC 800.0degC 0.07932U 56.005 5520 800degC _K TM N 2W 56.006 MATH MEM = MEM + (M[19]) 56.007 MATH MEM1 = 800 56.008 MEME 56.009 MEMC degC 0.5U 57.001 DOS METLINK3 Setup Source TC 1370 degC K 57.002 JMPZ 57.001 57.003 JMPF 103.001 57.004 ACC 1370.0degC 0.09566U 57.005 5520 1370.0degC _K TM N 2W 57.006 MATH MEM = MEM + (M[24]) 57.007 MATH MEM1 = 1370 57.008 MEME 57.009 MEMC degC 0.5U 58.001 DOS METLINK3 Off 58.002 JMPZ 58.001 58.003 JMPF 103.001 58.004 5520 * S 58.005 HEAD {DC CURRENT MEASUREMENT TEST}[7][7][7] 58.006 DISP The Fluke 5520A Current Output is characterized at 58.006 DISP the time of test and is displayed as the true value 58.006 DISP on the report of calibration. 58.007 HEAD DC CURRENT MEASUREMENT TEST {30mA Range} 58.008 PIC 58_74X_A 58.009 DOS METLINK3 Setup Measurement DCI 30E-3 58.010 JMPZ 58.009 58.011 JMPF 103.001 58.012 5520 4.000mA S 2W 58.013 3458 4.000mA N 2W 58.014 MATH M[1] = MEM 58.015 DOS METLINK3 Read DCI 58.016 JMPZ 58.009 58.017 JMPF 103.001 # ACC line reflects 3458A uncertainty. 58.018 ACC 4.000mA 4.9e-6U 58.019 MATH MEM1 = M[1] 58.020 MATH MEM = MEM * 1000 58.021 MEMC 30 mA 0.01% 0.015/ 59.001 5520 20.000mA S 2W 59.002 3458 20.000mA N 2W 59.003 MATH M[1] = MEM 59.004 DOS METLINK3 Read DCI 59.005 JMPZ 59.001 59.006 JMPF 103.001 # ACC line reflects 3458A uncertainty. 59.007 ACC 20.000mA 6.5e-6U 59.008 MATH MEM1 = M[1] 59.009 MATH MEM = MEM * 1000 59.010 MEMC 30 mA 0.01% 0.015/ 60.001 5520 30.000mA S 2W 60.002 3458 30.000mA N 2W 60.003 MATH M[1] = MEM 60.004 DOS METLINK3 Read DCI 60.005 JMPZ 60.001 60.006 JMPF 103.001 # ACC line reflects 3458A uncertainty. 60.007 ACC 30.000mA 7.5e-6U 60.008 MATH MEM1 = M[1] 60.009 MATH MEM = MEM * 1000 60.010 MEMC 30 mA 0.01% 0.015/ 61.001 5520 -30.000mA S 2W 61.002 3458 -30.000mA N 2W 61.003 MATH M[1] = MEM 61.004 DOS METLINK3 Read DCI 61.005 JMPZ 61.001 61.006 JMPF 103.001 # ACC line reflects 3458A uncertainty. 61.007 ACC -30.000mA 7.5e-6U 61.008 MATH MEM1 = M[1] 61.009 MATH MEM = MEM * 1000 61.010 MEMC 30 mA 0.01% 0.015/ 62.001 5520 * S 62.002 HEAD DC CURRENT MEASUREMENT: {110mA Range} 62.003 DOS METLINK3 Setup Measurement DCI 100E-3 62.004 JMPZ 62.003 62.005 JMPF 103.001 62.006 5520 0.00mA S 2W 62.007 3458 0.00mA N 2W 62.008 MATH M[1] = MEM 62.009 DOS METLINK3 Read DCI 62.010 JMPZ 62.003 62.011 JMPF 103.001 # ACC line reflects 3458A uncertainty. 62.012 ACC 0.00mA 1.65e-5U 62.013 MATH MEM1 = M[1] 62.014 MATH MEM = MEM * 1000 62.015 MEMC 110 mA 0.01% 0.015/ 63.001 5520 100.00mA S 2W 63.002 3458 100.00mA N 2W 63.003 MATH M[1] = MEM 63.004 DOS METLINK3 Read DCI 63.005 JMPZ 63.001 63.006 JMPF 103.001 # ACC line reflects 3458A uncertainty. 63.007 ACC 100.00mA 1.45e-5U 63.008 MATH MEM1 = M[1] 63.009 MATH MEM = MEM * 1000 63.010 MEMC 110 mA 0.01% 0.015/ 64.001 5520 -100.00mA S 2W 64.002 3458 -100.00mA N 2W 64.003 MATH M[1] = MEM 64.004 DOS METLINK3 Read DCI 64.005 JMPZ 64.001 64.006 JMPF 103.001 # ACC line reflects 3458A uncertainty. 64.007 ACC -100.00mA 1.45e-5U 64.008 MATH MEM1 = M[1] 64.009 MATH MEM = MEM * 1000 64.010 MEMC 110 mA 0.01% 0.015/ 65.001 5520 * S 65.002 HEAD {SIMULATE TRANSMITTER TEST}[7][D100][7][D100][7] 65.003 PIC 58_70xST 65.004 IEEE [@3458]DCI # Ignore 701 "Open Circuit Detected" error until 5520A output is applied. 65.005 DOS METLINK3 Setup Source Simulate_Transmitter 4E-3 Ignore 65.006 JMPZ 65.004 65.007 JMPF 103.001 65.008 5520 4V S 2W 65.009 DOS METLINK3 Setup Source Simulate_Transmitter 4E-3 65.010 JMPZ 65.004 65.011 JMPF 103.001 65.012 3458 22 4.0000mA 0.02% 0.03/ F 2W 66.001 IEEE [@3458]DCI 66.002 DOS METLINK3 Setup Source Simulate_Transmitter 22E-3 66.003 JMPZ 66.001 66.004 JMPF 103.001 66.005 5520 4V S 2W 66.006 3458 22 22.0000mA 0.02% 0.03/ F 2W 67.001 DOS METLINK3 Off 67.002 JMPZ 67.001 67.003 JMPF 103.001 67.004 5520 * S 67.005 HEAD {DC CURRENT SOURCING TEST}[7][D100][7][D100][7] 67.006 PIC 58_70x_A 67.007 IEEE [@3458]DCI 67.008 DOS METLINK3 Setup Source DCI 2E-3 67.009 JMPZ 67.001 67.010 JMPF 103.001 67.011 IEEE [D2000] 67.012 DOS METLINK3 Setup Source DCI 2E-3 67.013 JMPZ 67.007 67.014 JMPF 103.001 67.015 3458 22 2.0000mA 0.01% 0.015/ 2W 68.001 IEEE [@3458]DCI 68.002 DOS METLINK3 Setup Source DCI 4E-3 68.003 JMPZ 68.001 68.004 JMPF 103.001 68.005 IEEE [D1000] 68.006 DOS METLINK3 Setup Source DCI 4E-3 68.007 JMPZ 68.001 68.008 JMPF 103.001 68.009 3458 22 4.0000mA 0.01% 0.015/ 2W 69.001 IEEE [@3458]DCI 69.002 DOS METLINK3 Setup Source DCI 22E-3 69.003 JMPZ 69.001 69.004 JMPF 103.001 69.005 IEEE [D1000] 69.006 DOS METLINK3 Setup Source DCI 22E-3 69.007 JMPZ 69.001 69.008 JMPF 103.001 69.009 3458 22 22.0000mA 0.01% 0.015/ 2W 70.001 DOS METLINK3 Off 70.002 JMPZ 70.001 70.003 JMPF 103.001 70.004 HEAD {FREQUENCY SOURCING - SQUARE WAVE}[7][D100][7][D100][7] # System uncertainty of the PM6666 is inclusive of the instrument quantization # error,timebase error,systematic errors, and front end noise. 70.005 PIC 66_70xFR 70.006 DOS METLINK3 Setup Source Freq 5 1 70.007 JMPZ 70.006 70.008 JMPF 103.001 70.009 HEAD FREQUENCY SOURCING {Duty Cycle Test - 5Hz} 70.010 6666 * S 70.011 M666 COM 1T 0.5V 0.5V DC -+ X1 70.012 6666 100.00mT 1Vp TI N 70.013 MATH M[19] = MEM 70.014 6666 * S 70.015 M666 COM 1T 0.5V 0.5V DC +- X1 70.016 6666 100.00mT 1Vp TI N 70.017 MATH M[20] = MEM 70.018 MATH MEM = (M[20] / (M[19] + M[20])) * 100 70.019 ACC 50.0%ÿ 0.1% 70.020 MEME 70.021 MEMC %_ 2.5U 71.001 HEAD FREQUENCY SOURCING {Frequency Accuracy Test} 71.002 6666 5.00H 1Vp FA N 71.003 ACC 5.00H 4.5e-6U 71.004 MEME 71.005 MEMC H 0.01U 1Vp 72.001 DOS METLINK3 Setup Source Freq 50 1 72.002 JMPZ 72.001 72.003 JMPF 103.001 72.004 6666 50.0H 1Vp FA N 72.005 ACC 50.0H 4.5e-5U 72.006 MEME 72.007 MEMC H 0.1U 1Vp 73.001 DOS METLINK3 Setup Source Freq 500 1 73.002 JMPZ 73.001 73.003 JMPF 103.001 73.004 6666 500.0H 1Vp FA N 73.005 ACC 500.0H 4.50e-3U 73.006 MEME 73.007 MEMC H 0.1U 1Vp 74.001 DOS METLINK3 Setup Source Freq 10E+3 1 74.002 JMPZ 74.001 74.003 JMPF 103.001 74.004 6666 10.000kH 1Vp FA N 74.005 ACC 10.000kH 9e-6U 74.006 MEME 74.007 MEMC kH 0.002U 1Vp 75.001 DOS METLINK3 Setup Source Freq 50E+3 1 75.002 JMPZ 75.001 75.003 JMPF 103.001 75.004 6666 50.000kH 1Vp FA N 75.005 ACC 50.000kH 4.5e-5U 75.006 MEME 75.007 MEMC kH 0.005U 1Vp 76.001 DOS METLINK3 Setup Source FreqSine 1000 1 76.002 JMPZ 76.001 76.003 JMPF 103.001 76.004 M666 ChA 0.1V AC + X1 76.005 6666 1000.0H 1Vpp FA N 76.006 M666 * # Save results to include under SINE WAVE heading 76.007 MATH M[1] = MEM 76.008 MATH M[2] = MEM1 76.009 HEAD FREQUENCY SOURCING[7][7][7] 76.010 HEAD FREQUENCY SOURCING {Voltage Accuracy Test} 76.011 PIC 58_70x_V 76.012 DOS METLINK3 Setup Source Freq 5 1 76.013 JMPZ 76.012 76.014 JMPF 103.001 76.015 3458 0.500V 5H E N 2W 76.016 MATH MEM = MEM * 2 76.017 MEMC 10 1.000Vpp 3.0% 0.5/ 5H 77.001 DOS METLINK3 Setup Source Freq 50 1 77.002 JMPZ 77.001 77.003 JMPF 103.001 77.004 3458 0.500V 50H F N 2W 77.005 MATH MEM = MEM * 2 77.006 MEMC 10 1.000Vpp 3.0% 0.5/ 50H 78.001 DOS METLINK3 Setup Source Freq 500 1 78.002 JMPZ 78.001 78.003 JMPF 103.001 78.004 3458 0.500V 500H F N 2W 78.005 MATH MEM = MEM * 2 78.006 MEMC 10 1.000Vpp 3.0% 0.5/ 500H 79.001 DOS METLINK3 Setup Source Freq 10E+3 1 79.002 JMPZ 74.001 79.003 JMPF 103.001 79.004 3458 0.500V 10kH F N 2W 79.005 MATH MEM = MEM * 2 79.006 MEMC 10 1.000Vpp 10% 0.5/ 10kH 80.001 DOS METLINK3 Setup Source Freq 50E+3 1 80.002 JMPZ 80.001 80.003 JMPF 103.001 80.004 3458 0.500V 50kH F N 2W 80.005 MATH MEM = MEM * 2 80.006 MEMC 10 1.000Vpp 30% 0.5/ 50kH 81.001 HEAD {FREQUENCY SOURCING - SINE WAVE} # Evaluate results obtained earlier with PM6666 81.002 HEAD FREQUENCY SOURCING {Frequency Accuracy Test} 81.003 MATH MEM = M[1] 81.004 MATH MEM1 = M[2] 81.005 ACC 1000.0H 9.09e-4U 81.006 MEME 81.007 MEMC H 0.1U 1Vpp 82.001 HEAD FREQUENCY SOURCING {Voltage Accuracy Test} 82.002 DOS METLINK3 Setup Source FreqSine 1000 1 82.003 JMPZ 82.002 82.004 JMPF 103.001 82.005 3458 0.3535V 1000H F N 2W 82.006 MATH MEM = MEM * 2.828 82.007 MEMC 10 1.000Vpp 3% 0.5/ 1000H 83.001 DOS METLINK3 Off 83.002 JMPZ 83.001 83.003 JMPF 103.001 83.004 HEAD {DC VOLTAGE SOURCE} 83.005 HEAD DC VOLTAGE SOURCE: {110mV Range} 83.006 DOS METLINK3 Setup Source DCV -10E-3 83.007 JMPZ 83.001 83.008 JMPF 103.001 83.009 3458 110 -10.0000mV 0.01% 0.005/ F 2W 84.001 DOS METLINK3 Setup Source DCV 100E-3 84.002 JMPZ 84.001 84.003 JMPF 103.001 84.004 3458 110 100.0000mV 0.01% 0.005/ F 2W 85.001 HEAD DC VOLTAGE SOURCE: {1.1V Range} 85.002 DOS METLINK3 Setup Source DCV 0.12 85.003 JMPZ 85.001 85.004 JMPF 103.001 85.005 3458 1.1 0.120000V 0.01% 0.005/ F 2W 86.001 DOS METLINK3 Setup Source DCV 1 86.002 JMPZ 86.001 86.003 JMPF 103.001 86.004 3458 1.1 1.000000V 0.01% 0.005/ F 2W 87.001 HEAD DC VOLTAGE SOURCE: {15V Range} 87.002 DOS METLINK3 Setup Source DCV 1.2 87.003 JMPZ 87.001 87.004 JMPF 103.001 87.005 3458 15 1.20000V 0.01% 0.005/ F 2W 88.001 DOS METLINK3 Setup Source DCV 14 88.002 JMPZ 88.001 88.003 JMPF 103.001 88.004 3458 15 14.00000V 0.01% 0.005/ F 2W 89.001 DOS METLINK3 Off 89.002 JMPZ 89.001 89.003 JMPF 103.001 89.004 HEAD {RESISTANCE SOURCING TEST}[7][D100][7][D100][7] 89.005 PIC 58_70xCW 89.006 HEAD RESISTANCE SOURCING TEST {11 Ohm Range} 89.007 M3458 RNGLK 100Z 89.008 DOS METLINK3 Setup Source Ohms 0.1 89.009 JMPZ 89.001 89.010 JMPF 103.001 89.011 3458 11 0.1000Z 0.01% 0.02U O F 4W 90.001 DOS METLINK3 Setup Source Ohms 1 90.002 JMPZ 90.001 90.003 JMPF 103.001 90.004 3458 11 1.0000Z 0.01% 0.02U O F 4W 91.001 DOS METLINK3 Setup Source Ohms 10 91.002 JMPZ 91.001 91.003 JMPF 103.001 91.004 3458 11 10.0000Z 0.01% 0.02U O F 4W 92.001 M3458 * 92.002 HEAD RESISTANCE SOURCE: {110 Ohm Range} 92.003 DOS METLINK3 Setup Source Ohms 20 92.004 JMPZ 92.002 92.005 JMPF 103.001 92.006 3458 110 20.000Z 0.01% 0.04U O F 4W 93.001 DOS METLINK3 Setup Source Ohms 100 93.002 JMPZ 93.001 93.003 JMPF 103.001 93.004 3458 110 100.000Z 0.01% 0.04U O F 4W 94.001 HEAD RESISTANCE SOURCE: {1100 Ohm Range} 94.002 DOS METLINK3 Setup Source Ohms 0.2E+3 94.003 JMPZ 94.001 94.004 JMPF 103.001 94.005 3458 1.1 200.00Z 0.02% 0.5U O F 4W 95.001 DOS METLINK3 Setup Source Ohms 1E+3 95.002 JMPZ 95.001 95.003 JMPF 103.001 95.004 3458 1.1 1000.00Z 0.02% 0.5U O F 4W 96.001 HEAD RESISTANCE SOURCE: {11 kOhm Range} 96.002 DOS METLINK3 Setup Source Ohms 2E+3 96.003 JMPZ 96.001 96.004 JMPF 103.001 96.005 3458 11 2.0000kZ 0.03% 0.005U O F 4W 97.001 DOS METLINK3 Setup Source Ohms 10E+3 97.002 JMPZ 97.001 97.003 JMPF 103.001 97.004 3458 11 10.0000kZ 0.03% 0.005U O F 4W 98.001 DOS METLINK3 Off 98.002 JMPZ 98.001 98.003 JMPF 103.001 98.004 HEAD {RTD SOURCE TEST} 98.005 DOS METLINK3 Setup Source RTD -180 degC PT385 98.006 JMPZ 98.001 98.007 JMPF 103.001 98.008 3458 27.096Z 0.043U O F 4W 99.001 DOS METLINK3 Setup Source RTD 100 degC PT385 99.002 JMPZ 99.001 99.003 JMPF 103.001 99.004 3458 138.505Z 0.076U O F 4W 100.001 DOS METLINK3 Setup Source RTD 780 degC PT385 100.002 JMPZ 100.001 100.003 JMPF 103.001 100.004 3458 369.712Z 0.12U O F 4W 101.001 DOS METLINK3 Off 101.002 JMPZ 101.001 101.003 JMPF 103.001 101.004 HEAD {LOOP POWER TEST}[7][D100][7][D100][7][D100][7] 101.005 PIC 58_70xLP 101.006 DOS METLINK3 Setup Source Loop_Pwr 24 101.007 JMPZ 101.001 101.008 JMPF 103.001 101.009 3458 24.0V 1.2U F 2W 102.001 DOS METLINK3 Setup Source Loop_Pwr 28 102.002 JMPZ 102.001 102.003 JMPF 103.001 102.004 3458 28.0V 1.4U F 2W 103.001 DOS METLINK3 Off 103.002 JMPZ 103.001 103.003 END